Immunity test
Test / Compliance level
Electromagnetic environment – guidance
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 0% U
T
for 0,5 cycle at
45° steps
0% U
T
for 1 cycles
70% U
T
for 25 / 30 cycles
< 5% U
T
for 250 / 300 cycles
Mains power quality should be that of a typical commercial or hospital environment.
If the user of this product requires continued operation during power mains interrup-
tions, it is recommended that the product is powered from an un-interruptible power
supply or a battery.
U
T
is the a. c. mains voltage prior to application of the test level.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
30 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location
in a typical commercial or hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3V
150 kHz to 80 Mhz
6V
in ISM & amateur radio
bands
10 V/m
80 Mhz to 2,7 GHz
385 MHz - 5785 MHz test
specifications
for immunity to RF wireless
communication equipment
refer to table 9 of IEC
60601-1-2:2014
Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the elec-
tromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which this prod-
uct is used exceeds the applicable RF compliance level above, this product should be
observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating this product.
Interference may occur in the vicinity of equipment marked with the
following symbol:
Portable and mobile RF communications equipment should be used no closer than 30
cm to any part of this product including cables.
Test Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
power (W)
Distance (m)
Immunity test level
(V/m)
385
380 - 390
TETRA 400
Pulse modulation
b)
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c)
± 5 kHz
deviation
1 kHz sine
2
0.3
28
710
745
780
704 - 780
LTE Band 13,17
Pulse modulation
b)
217 Hz
0.2
0.3
9
810
870
930
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850, LTE
Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
10.
3.1 Test specifications for immunity to RF wireless communications equipment
IEC 60601-1-2:2014 — Table 9