14E. Min/Max:
Toggles the numerical display to only record a value that in Min, Is less than the
exis ng value on the display or in Max is greater than the exis ng value on the display.
14E.1 Min/Max:
also doubles as a
Peak FW
bu3on. Peak FW is similar to Max, except it pushes
the limits of the system’s rapid measurement capability and uses the firmware speed to scan for
changes that occur very fast.
14F. 100%:
Records the baseline value for 100%. This does not impact the display. To readout in
percentage you must use the display units bu3on to select percentage.
14G. Record:
Toggles the record data to on and off. When on, the meter stores all readings internally.
When off the meter stops saving readings to the internal memory. Recorded measurements generate
a new file each me the record start bu3on is touched. The format for naming recorded files is ILT
day-mon-year. Record creates a five column .csv file with Time, Instantaneous, min, max and average
results.
14H. Capture
: Records a single reading at the me the capture bu3on is pressed. Captured
measurements are appended to the single sample file, in the format date/ me, instantaneous value.
14I. Home
Returns to the home screen.
15 Se5ngs:
The settings icons open the settings display which allow user customization. Settings can
also be access by clicking on the setting drop down menu in the dis
play
screens.
15A. Display Brightness
: Brightness selec on ranges from 1-Dim with lowest ba3ery consump on, to
10-Brightest with highest ba3ery consump on.
Radiometer Se$ngs
15B Calibra on factors:
15B1. Meter Factor :
Selects calibra on factors from the pre-programmed/available list ranging from
1 – 20. Factor selec on allows readings in calibrated units such as W, W/cm2, Lux etc.
15B2. Smart Detector Factor:
Smart detectors can also store up to 20 calibra on factors. In order to
use, or create cal factors that are saved in the sensor connector, Use Smart Detector Factor must be
enabled, in 15B3.
15Be. USE Smart Detector Factor: ON/OFF
ON enable the use of the sensors cal factors. Off only
allows customer to use factors save in the ILT2500 meter.
15C. Sample Time:
Allows the user to program how fast the meter takes a reading ranging from 100
milliseconds to 5 seconds with the default set to Auto.
Auto will review the incoming signals intensity and apply the best seLngs. (faster with higher light
levels, slower for smaller signals for be3er averaging)As a general rule of thumb, If current >= 1uA,
0.5sec is used, else if current <1uA, 1.0sec
FAST samples at 20ms (50Hz)
Note: Using trend at fast speeds taxes the CPU with processing both measurements, numerical
updates and graph updates. This can be seen as a slight flicker on the display. If this problem occurs,
use Meter (no graph) or connect to the computer and use Datalight III Meter or Flash.
15D. Auto Gain Range:
On or off. The ILT2500 has 3 relays that it cycles through when auto ranging
from the lowest signals in the Pico amp range to the higher signals in the micro amp to 1 milli amp
range. Auto allows the so ware to control when the device will change from one relay to the next to
allow the most accurate results based on incoming current levels. Auto gain range- off requires the
ILT2500 INSTRUMENTATION MANUAL Preliminary 8/18 11 OF 22