compa bility with ILT1700 (and IL1350) sensors. Future ILT2500 sensors will be supplied with an 
internal memory chip inside the D-sub connector that includes the model and serial number of the 
sensor as well as calibra on informa on. The ILT2500 will have the ability to read the smart detector 
informa on and apply the proper calibra on factor to be used during tes ng.  
The D-sub not only receives the current generated by the sensor a er exposure to light, it also 
supplies  bias to sensors requiring bias for power and/or to improve response  me.  The D sub 
connector is not limited to only measuring devices manufactured by ILT. Any solid state device that 
generates current from 50 pA to 1 mA can be connected to the ILT2500 for measurement of current. 
Please see wiring diagram below:

Wiring diagram for D sub:

PinDe

scrip 

on

Pin

Descri

p on

Pin1:

-9Volts

Pin9:

Shield

Pin2:

I2C SCL

Pin10: SODIMM_85

Pin3:

I2C SDA

Pin11: SODIMM_97

Pin4:

3.3Volts

Pin12: PWM_B

Pin5:

GND

Pin13: External Analog In 1

Pin6:

Det. Cathode

Pin14: External Analog In 2

Pin7:

Det. Anode

Pin15: N/C

Pin8:

5Volts

Warning: please consult with an ILT technician before connec ng a current genera ng device that 
could exceed 1 mA of current to prevent risk of damage to the sensi ve measurement circuitry. 

6. USB Ports: 

There are three USB ports on the side of the

ILT2500 device

UPDATE:

 Is used for code updates and requires the use of

a memory s ck. (not included) 

DATA:

  Accesses the internal 4 Gb memory device where data is saved and to charge the 

ba3ery. The Micro USB is used to download stored data and to erase data from the memory 
device.  This port can also be used the charge the ba3ery.  

PC:

  The Mini USB port is used to connect to a computer and to charge the ba3ery.   Datalight 

so ware is available for download on the website: h3ps://www.intl-
ligh3ech.com/support/so ware, or request an emailed copy by emailing ilsales@intl-
ligh3ech.com

7. Display:

The Display is a full color

 4.3" Capaci ve TFT Touch screen display.  Contact with sharp objects will 

damage the display. A stylus or finger is recommended for changing seLngs on the display. Care 

ILT2500 INSTRUMENTATION MANUAL  Preliminary 8/18     7 OF 22

Summary of Contents for ILT2500

Page 1: ...ILT2500 Opera on Manual ILT2500 manual www intl lighttech com 1 of 22...

Page 2: ...t Measurement and Limita ons 8 9 Sample Time and Limita ons 9 10 5V Bias 9 11 Home Screen 9 12 Display Style Selec on 10 13 Display Content 10 14 Feature Bu ons 11 15 Se5ngs 11 16 Files 16 17 Light Ca...

Page 3: ...480x272 TFT Color LCD Touch enabled Backlit C Opera ng Temperature 0 40 Deg C D Storage Temperature 10 45 Deg C E Memory 4GB microSD F USB Current Draw Typical 500mA Max 900mA USB 3 0 G Bias Programm...

Page 4: ...in sensors mfg 2018 or later 2 Turn on the meter by pushing the On O bu3on down into the I posi on 3 A er a short start up menu the Home Screen will appear 4 Check the power monitor window in the upp...

Page 5: ...mbient zero is used to subtract small background light levels that are not coming from the test source With the test source o or blocked and the sensor exposed to the environment ambient light press z...

Page 6: ...al temperature can rapidly increase and cause erroneous readings and damage Standard opera ng temperature is rated at 0 400 C Do not apply any force to the display Do not apply pressure to extended st...

Page 7: ...Analog In 1 Pin6 Det Cathode Pin14 External Analog In 2 Pin7 Det Anode Pin15 N C Pin8 5Volts Warning please consult with an ILT technician before connec ng a current genera ng device that could excee...

Page 8: ...hen this condi on occurs the display will provide a SATURATION warning To prevent satura on neutral density lters or apertures can be added to the sensor or a greater distance between the sensor and t...

Page 9: ...material ILT silicon sensors include SED033 SED100 SCD110 XSD140B XSD340B XSD140A and XSD340A 11 Home Screen When the meter is turned on it will take a few moments for the so8ware to ini alize Upon c...

Page 10: ...e as fast as 1 seconds or as slow as once per hour Note See see 15I 15J for seLng bar or line graph 14 Feature Bu ons 14A Display Units Toggles between the available fields Current output is in amps O...

Page 11: ...alibra on factors 15B1 Meter Factor Selects calibra on factors from the pre programmed available list ranging from 1 20 Factor selec on allows readings in calibrated units such as W W cm2 Lux etc 15B2...

Page 12: ...es such as scanning lasers or pulses of light use FW peak For be3er averaging we recommend using min max when speed is not a factor 15G 5V Bias On or O The defaults seLng is o as there are numerous se...

Page 13: ...3 For High and Range the error level must be greater than the warning level Ie target is 1 mW cm2 1e 3 the warning must be larger than 1e 3 ie 3e 3 or 3 mW and the error must be larger than the warni...

Page 14: ...ed arrow on cal cert on page 14 15Q 5 Resul ng Light Level Allows the user to enter a know value for the light level under test and create a calibra on factor to force the meter to read the correct va...

Page 15: ...adings When the ILT2500 is connected to the computer using the USB micro cable the computer will register the ILT2500 as a device with removable storage allowing access into the Files folder To access...

Page 16: ...the loca on to apply With the box showing the loca on touch the Del delete bu3on The Delete bu3ons border will turn blue to indicate comple on If you want to back out of Delete scroll to the top and...

Page 17: ...in SeLngs The white display box shows the ash count in the upper le Note Counts was designed for rapid rises in light intensity and increases when the signal rises 4x between 100uS samples the working...

Page 18: ...generated light level measured exceeds the measurement capabili es of the system Increase the distance add a3enua on lters or add an aperture to the sensor to allow measurement of stronger signals Low...

Page 19: ...00 decreases the number of digits of strobes to count v is a programmable value of how many ashes to be measured Level selec on High Light level 3 uA to 1 mA Frequency 3 1 4 Hz Low Light 0 to 3 uA Fre...

Page 20: ...hould be plugged in during ini aliza on Insert the USB s ck into the USB A port on the side of the ILT2500 Click on Settings and Scroll to the bottom of setting window Click on the rmware update box S...

Page 21: ...pressure to the meter when the stand is in use 28A Opera ng Temperature 0 40 C ILT meters and sensors have a recommended opera ng range of 0 to 40 C to assure the accuracy of the calibrated measuremen...

Page 22: ...parts or repair by other than Interna onal Light Technologies The equipment will be repaired or replaced at our op on without charge to the owner for parts or labor incurred in such repair This warra...

Reviews: