3F. Quick Start – Common Problems:

 The two most common causes of error are incorrect calibra on 

factor use and improper zero. It is a good prac ce to verify the correct calibra on factor is in use 
before taking measurements.  It is equally as important to assure a proper zero has been performed. 

3F1. Calibra on Factor Verifica on.

 Locate the calibra on factor on the detector calibra on 

cer ficate below the model and serial number fields in the top third of the report.
Example: 

From the home screen select seLngs, scroll to the second seLng window labeled Factor Number. 
Verify the sensi vity factor and factor units are correct. If not touch the selec on window and a 
scroll menu will appear, select a new channel un l you find the correct sensi vity factor as noted 
on your cer ficate.  (If you do not locate your calibra on factor and units from the cer ficate, go 
to 5D to learn about adding/changing calibra on factors)

  Touch the back bu3on to return to the home screen. 

3F2. Zero:

 There are two kinds of zero that can be used to improve accuracy. A 

dark zero

 

is used 

to remove small ambient noise caused from the environment. With the sensor plugged in and 
covered with an opaque object, press the zero bu3on. This reading will be subtracted from all 
future readings to provide a zero reading.  An 

ambient zero

 is used to subtract small background 

light levels that are not coming from the test source. With the test source off or blocked and the 
sensor exposed to the environment/ambient light, press zero.  (An example of an ambient zero 
source is a computer display that is needed during tes ng.) Place the un-covered sensor in the 
test loca on. Cover the test light source, or turn it off. Press Zero. The ILT2500 will take a 
measurement and subtract that value from all future readings un l zero is pressed again.

3G. Quick Start Feature Overview:

 The first  me you turn on your meter you will most likely open in 

the home screen. IF not, you can use the home bu3on to access the home scrren. Select an 
applica on such as Meter or Trend to familiarize yourself with the func onality.  The App you are 
running (ie Meter Tall) will be displayed in the upper le  and the ba3ery life indicator will be displayed
in the upper right.   In the second line you will see the units of measurement and the selected 
calibra on factor Channel. 

3G1. Display resolu on: 

 The < - and- > bu3ons below the display will allow the user the control 

how many digits are displayed. Ie. 0.11 changes to 0.1111.

3G2. Display Units:

 Toggles between the available fields; current, light level, OD, and % trans.

To take calibrated light level readings, press display units un l light level is selected.

Note: 

If light level is selected prior to selec ng a calibra on/sensi vity factor, an error 

ILT2500 INSTRUMENTATION MANUAL  Preliminary 8/18     5 OF 22

Summary of Contents for ILT2500

Page 1: ...ILT2500 Opera on Manual ILT2500 manual www intl lighttech com 1 of 22...

Page 2: ...t Measurement and Limita ons 8 9 Sample Time and Limita ons 9 10 5V Bias 9 11 Home Screen 9 12 Display Style Selec on 10 13 Display Content 10 14 Feature Bu ons 11 15 Se5ngs 11 16 Files 16 17 Light Ca...

Page 3: ...480x272 TFT Color LCD Touch enabled Backlit C Opera ng Temperature 0 40 Deg C D Storage Temperature 10 45 Deg C E Memory 4GB microSD F USB Current Draw Typical 500mA Max 900mA USB 3 0 G Bias Programm...

Page 4: ...in sensors mfg 2018 or later 2 Turn on the meter by pushing the On O bu3on down into the I posi on 3 A er a short start up menu the Home Screen will appear 4 Check the power monitor window in the upp...

Page 5: ...mbient zero is used to subtract small background light levels that are not coming from the test source With the test source o or blocked and the sensor exposed to the environment ambient light press z...

Page 6: ...al temperature can rapidly increase and cause erroneous readings and damage Standard opera ng temperature is rated at 0 400 C Do not apply any force to the display Do not apply pressure to extended st...

Page 7: ...Analog In 1 Pin6 Det Cathode Pin14 External Analog In 2 Pin7 Det Anode Pin15 N C Pin8 5Volts Warning please consult with an ILT technician before connec ng a current genera ng device that could excee...

Page 8: ...hen this condi on occurs the display will provide a SATURATION warning To prevent satura on neutral density lters or apertures can be added to the sensor or a greater distance between the sensor and t...

Page 9: ...material ILT silicon sensors include SED033 SED100 SCD110 XSD140B XSD340B XSD140A and XSD340A 11 Home Screen When the meter is turned on it will take a few moments for the so8ware to ini alize Upon c...

Page 10: ...e as fast as 1 seconds or as slow as once per hour Note See see 15I 15J for seLng bar or line graph 14 Feature Bu ons 14A Display Units Toggles between the available fields Current output is in amps O...

Page 11: ...alibra on factors 15B1 Meter Factor Selects calibra on factors from the pre programmed available list ranging from 1 20 Factor selec on allows readings in calibrated units such as W W cm2 Lux etc 15B2...

Page 12: ...es such as scanning lasers or pulses of light use FW peak For be3er averaging we recommend using min max when speed is not a factor 15G 5V Bias On or O The defaults seLng is o as there are numerous se...

Page 13: ...3 For High and Range the error level must be greater than the warning level Ie target is 1 mW cm2 1e 3 the warning must be larger than 1e 3 ie 3e 3 or 3 mW and the error must be larger than the warni...

Page 14: ...ed arrow on cal cert on page 14 15Q 5 Resul ng Light Level Allows the user to enter a know value for the light level under test and create a calibra on factor to force the meter to read the correct va...

Page 15: ...adings When the ILT2500 is connected to the computer using the USB micro cable the computer will register the ILT2500 as a device with removable storage allowing access into the Files folder To access...

Page 16: ...the loca on to apply With the box showing the loca on touch the Del delete bu3on The Delete bu3ons border will turn blue to indicate comple on If you want to back out of Delete scroll to the top and...

Page 17: ...in SeLngs The white display box shows the ash count in the upper le Note Counts was designed for rapid rises in light intensity and increases when the signal rises 4x between 100uS samples the working...

Page 18: ...generated light level measured exceeds the measurement capabili es of the system Increase the distance add a3enua on lters or add an aperture to the sensor to allow measurement of stronger signals Low...

Page 19: ...00 decreases the number of digits of strobes to count v is a programmable value of how many ashes to be measured Level selec on High Light level 3 uA to 1 mA Frequency 3 1 4 Hz Low Light 0 to 3 uA Fre...

Page 20: ...hould be plugged in during ini aliza on Insert the USB s ck into the USB A port on the side of the ILT2500 Click on Settings and Scroll to the bottom of setting window Click on the rmware update box S...

Page 21: ...pressure to the meter when the stand is in use 28A Opera ng Temperature 0 40 C ILT meters and sensors have a recommended opera ng range of 0 to 40 C to assure the accuracy of the calibrated measuremen...

Page 22: ...parts or repair by other than Interna onal Light Technologies The equipment will be repaired or replaced at our op on without charge to the owner for parts or labor incurred in such repair This warra...

Reviews: