EN – English
18
2. Grasp the Adjustment Knob and pull it away from the Locking Bracket to unlock.
3. Holding the Adjustment Knob in its unlocked position, rotate it in the desired direction to
adjust tension. Clockwise (+ plus) direction to increase tension, or Counterclockwise (- minus)
direction to decrease tension.
Adjustment Knob
Increase Locking
Lever Tension
Decrease Locking
Lever Tension
Figure 10 Adjustment Knob
4. Test the operation of the handle. With the Locking Lever fully opened, the Locking Bracket
should rotate freely. With the Locking Lever closed, the Locking Bracket should not rotate.
NOTE: Once the desired setting is achieved confirm the Adjustment Knob is in the locked, seat-
ed position.
CAUTION!
It is possible to adjust the handle to the point that the Locking Lever requires
excessive force to close. Do not exert excessive force as this may result in damage
to the device if the lever is forced closed.