Disinfection
Further disinfection of device components may be achieved using one of the following methods.
Method 1:
Chemical
•
With device cleaned and still disassembled, wipe all surfaces with sterile gauze moistened with
70% Isopropyl Alcohol (IPA). Be sure to achieve even coverage on all surfaces.
•
Assure the device stays wet with the 70% IPA for a minimum of 10 minutes.
•
Dry the device with a sterile lint free cloth or blow dry with filtered pressurized air.
Method 2:
Thermal Rinse
•
A Thermal Disinfection phase maybe added after the rinse cycle as indicated in table 1.
Maintenance and Care:
The Infinity XR2 Base Unit should be returned to its storage case after each use.
NOTE: Ensure all components have completely dried after cleaning before returning to case.
CAUTION!
If unit is dropped or mishandled, it should be inspected for damage.
(REF Inspection section of this instruction manual) If damage occurs, do not use;
return the complete device immediately to Integra LifeSciences for inspection.
The life expectancy of the MAYFIELD products is expected to be as long as 7 years.
To ensure proper function and to extend the life and performance of the equipment, Integra
LifeSciences recommends the following:
Recommended Action
Recommended Frequency
Return the device to the Integra LifeSciences Repairs
department for detailed inspection and servicing.
Once / year
Request that Integra NeuroSpecialists perform routine
inspections of the device.
Twice / year
In the absence of proper care and servicing of the device, negative effects may be seen after repeated
processing over time which may lead to reduced performance.
Contact information: See the Service and Repair section for contact information on how to return your
device for periodic servicing and to request periodic inspections.
See Inspection and /or Service notes section for routine checks to be performed on the device.
NOTE: Any serious incident that has occurred in relation to the device for the user and/or the patient
should be reported to the manufacturer and the competent authority of the member state in which the
user and/or patient is established.
Device Disposal
NOTE : Follow hospital procedures for disposal of this device.
Summary of Contents for MAYFIELD Infinity XR2 A2079
Page 2: ......
Page 23: ...21 Unités de base MAYFIELD Infinity XR2 A2079 A2079E Mode d emploi ...
Page 26: ...FR Français 24 Figure 1A Composants de l unité de base A2079 Infinity XR2 ...
Page 43: ...41 Unità di Base Infinity XR2 di MAYFIELD A2079 A2079E Manuale di istruzioni ...
Page 63: ...61 MAYFIELD Infinity XR2 Grundeinheiten A2079 A2079E Gebrauchsanleitung ...
Page 83: ...81 Unidades base Infinity XR2 de MAYFIELD A2079 A2079E Manual de instrucciones ...
Page 103: ...101 MAYFIELD Infinity XR2 basisunits A2079 A2079E Gebruikershandleiding ...