EN – English
14
Optional Automatic Wash Cycle:
CAUTION!
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
•
Disassemble the device as indicated in section
Cleaning of the MAYFIELD Infinity XR2 Base
Unit.
•
Rinse components under warm tap water prior to placing in washer.
•
Arrange device in a way to avoid contact between components.
•
Select instrument cycle and ensure proper programming.
Table 1 Instrument Cycle
Phase
Time in minutes
Water
Temperature
Detergent type &
Concentration
Pre-Wash 1
04:00
Cold tap H
2
O
N/A
* Hot Tap water not to exceed 82.2C
** Optional phase for disinfection of components
Thermal Rinse**
Rinse 1
Wash 1
Enzyme Wash
04:00
10:00
00:30
02:00
Hot tap H
2
O*
60° C
Hot tap H
2
O*
82.2° C
Endozime® AW
Triple plus w/
A.P.A., 1:128 ratio
Renu-Klenz™,
1:256 ratio
N/A
N/A
Table 1 Instrument Cycle
•
Remove from washer and dry completely if needed.
•
Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible debris or residue is detected.
Summary of Contents for MAYFIELD Infinity XR2 A2079
Page 2: ......
Page 23: ...21 Unités de base MAYFIELD Infinity XR2 A2079 A2079E Mode d emploi ...
Page 26: ...FR Français 24 Figure 1A Composants de l unité de base A2079 Infinity XR2 ...
Page 43: ...41 Unità di Base Infinity XR2 di MAYFIELD A2079 A2079E Manuale di istruzioni ...
Page 63: ...61 MAYFIELD Infinity XR2 Grundeinheiten A2079 A2079E Gebrauchsanleitung ...
Page 83: ...81 Unidades base Infinity XR2 de MAYFIELD A2079 A2079E Manual de instrucciones ...
Page 103: ...101 MAYFIELD Infinity XR2 basisunits A2079 A2079E Gebruikershandleiding ...