9.1.3 Further disassembly of the top part – HeartScreen 60G:
1.
Disconnect the connection at the power button and the receptacles on the board accepting the printer head cables.
2.
Remove the Interface Board holding screws and remove it from the device (during re-assembly note the LCD
position: tilt, etc).
3.
To remove the LCD, the LCD has to be unsoldered from the pins connecting it to the Interface Board.
4.
Remove the LCD from the Interface Board by unscrewing the two spacer fixed screws.
5.
After removing the LCD, the keyboard can be pushed out from the device’s top part.
Interface board fixing screws
On/Off switch