2014.01.09 40/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
9.1.1 Further disassembly of the bottom part:
1. Remove the battery connections.
2.
If this was the goal of disassembly, remove the battery.
3.
After removing the indicated screws, the HS80G-L Main Board can be removed.
4.
The connecting cable of the insulating-shielding plate fixed to the tray is fixed with the top left card fixing screw,
as indicated on the diagram.
When re-assembling, make sure this cable is properly fixed.
5. The shielding fixed to the tray is glued from the bottom, and if removed, should be glued back.
Card fixing screws
Amplifier insulation
Card fixing screws
Card fixing screws