Innomed Medical
Heart Screen 60G-80G-L
Service manual
2014.01.09 38/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
Short screw
Short screw
Long screw
Long screw
Long screw
Connection label on the rear
9 Maintenance and service procedures
9.1
Device disassembly / re-assembly
The device has to be disassembled in the following order (re-assembly must be performed in reverse order):
1.
Remove the self-adhesive connection-label from the back (make sure that the label is not damaged, as it can be
replaced after re-assembly).
2. Turn the device upside-down and remove the five screws (during re-assembly take care to note the different length
screws).