Location of the reed-tube
9.2
Reprogramming the device
Beside the device, the following are required to reprogram the device:
1.
Serial cable
2.
PC
3.
Magnet for reprogramming
4.
The software update program
The reprogramming procedure
1.
Connect to the PC with the serial cable.
2.
Find the magnet’s proper position:
The magnet is required to operate the reed-tube at the backside of the device next to the charging adapter.
Closing the sensors in this places the device into the programming mode (the magnet can be anything
magnetic, for example something that magnetizes a screwdriver).
How to find it:
2.1.
The device is turned off; the serial cable is connected.
2.2.
Place the magnet tight against the device’s back, next to the charging adapter connection and turn the
device on.
If the reed-tube switched, nothing is visible on the display after it has been turned on. The device is in
programming mode. Remove the magnet and turning the device off and on again starts the device in
normal mode.
2.3.
If the tube did not switch, try different positions, moving the magnet polarity around the charging
adapter connection.
2.4.
After several unsuccessful tries read the Troubleshooting section of the manual.
3.
With the device turned on and the magnet in place, start the PC-based download program and follow the
instructions (magnet placement / removal, device on / off, etc).
4.
In case of incorrect download, repeat the process.
5.
After several unsuccessful tries, read the Troubleshooting section of the manual.