Innomed Medical
Heart Screen 60G-80G-L
Service manual
2014.01.09 6/84
D-5450-1-B_HS60G80GL_SM_ENG.doc
4.4
HeartScreen 80G-L operating modes
Monitoring
3-channel ECG monitoring on the LCD from any lead, including
SPEC group as well, plus 1 rhythm channel.
Within this:
Manual printing
Manual recording from any lead group on 3 channels, including
SPEC group.
Automatic printing
Adjustable print length from the 12 standard ECG leads in 4x3
channel groups.
Record storage
(STORE)
Automatic recording and storage from all groups, including SPEC
group.
Rhythm curve may also be stored in pre-programmed mode, on
which the device performs rhythm analysis as well.
Diagnosis creation
(DIAG)
Automatic 12-channel recording with analysis and diagnosis. Raw
curve, rhythm curve and SPEC group may also be stored in pre-
programmed mode.
Rhythm analysis
creation
(ARTM)
Rhythm analysis creation with statistics from the lead-group for the
selected lead.
Using the menus
Handling records stored in the device and settings.
Within this:
Catalog
(Menu / Catalog)
Used to handle stored records and related data.
Settings
(Menu / Settings)
Used to set basic device settings.
Identifier
(Menu / Identifier)
Used to enter numerical patient identifier.
4.5
HeartScreen 80G-L lead groups
Standard
3 channel: I-III, aVR-aVF, V1-V3, V4-V6
Cabrera
3-channel: aVL, I, aVR; II, aVF, III; V1-V3; V4-V6
SPEC lead group
first
second
third
any lead from the standard 12
any lead from the standard 12
any lead from the standard 12
A lead may only appear once in the spec group. Leads part of the standard groups appear normally on the
display, while leads part of the SPEC lead group appear in inverse.