Use
r’s
Manual of OMNI Express Patient Monitor
- 77
037.007.0001 Rev.2
Use this setting to fill in the bottom portion of the waveform on any
channel of the display.
SWEEP SPEED
12.5 mm/s and 25 mm/s are for choice, the factory setting is 12.5 mm/s.
SWEEP COLOR
It provides white, red, green, cyan, blue, yellow, gray and magenta which can be chosen.
GAS UNIT SETUP
“mmHg”, “kPa” and “%” are for choice.
ALARM SETUP
High
Low
FI Agt
0.0%-20.0%
factory setting: 5.0%
0.0%-20.0%
factory setting: 0.0%
ET Agt
0.0%-20.0%
factory setting: 5.0%
0.0%-20.0%
factory setting: 0.0%
FI CO2
0 mmHg-76 mmHg
factory setting: 4 mmHg
0 mmHg-76 mmHg
factory setting: 0 mmHg
ET CO2
0 mmHg-76 mmHg
factory setting: 61 mmHg
0 mmHg-76 mmHg
factory setting: 15 mmHg
RR
0 rpm-100 rpm
factory setting:
30 rpm (Adult
or Pediatric)
0 rpm-100 rpm
factory setting:
5 rpm (Adult or
Pediatric)
100
rpm
(Neonatal)
30
rpm
(Neonatal)
FI N2O
0%-100%
factory setting: 100%
0%-100%
factory setting: 0%
ET N2O
0%-100%
factory setting: 100%
0%-100%
factory setting: 0%
FI O2
18%-100%
factory setting: 100%
18%-100%
factory setting: 18%
ET O2
0%-100%
factory setting: 100%
0%-100%
factory setting: 5%
ADVANCED SETUP
ZERO GAS TYPE
“Scrubbed Air/N
2
/O
2
”, “Room Air” and “100% O
2
” for choice, the factory setting is “Room
Air”.
O2 COMPENSATIONS
The anesthetic agents in the gas mixture need to be compensated in order to achieve its
stated accuracy. The instrument settings for this parameter should be set when the O
2
sensor is unconnected. But when install the O
2
sensor, this function isn’t available.
STANDBY MODE
When measurements are temporarily not needed, the monitor can switch the AAM from
‘Operation Mode’ into ‘Standby Mode’. During standby, some internal components of the
module are switched off, which increases the lifetime of the module. When measurements
are needed again, the device must be switched back into
‘Operation Mode’. The latter
transition will usually take less than 30 seconds.