Resolution:
1 mmHg
Pressure Accuracy:
Maximum Mean Error:
5 mmHg
Maximum Standard Deviation: 8 mmHg
Overpressure Protection:
Adult/Pediatric Mode :
297 (mmHg)
Neonatal Mode :
147 (mmHg)
Alarm Limit:
SYS(Upper/Lower): 30
~
240 mmHg
DIA (Upper/Lower) : 15
~
180 mmHg
SpO2
Standard:
ISO 9919
ASpO2:
Anti-motion SpO
2
Measurement Technology:
Light absorption method
SpO
2
Probe:
Red Light LED Wavelength: 660±5 nm
Infrared Light LED Wavelength: 940±10 nm
Option Type:
BCI, Masimo, Nellcor
(See
each
module
’s
relative
technical
specifications)
Refresh Rate:
1 s
SpO
2
Alarm Limit:
Upper Limit : 50
~
100 %,
Lower Limit : 50
~
100 %
PR Alarm Limit:
Upper Limit : 70
~
239 bpm,
Lower Limit : 20
~
150 bpm
BCI SpO2
SpO
2
Measurement Range:
0
~
100 %
SpO
2
Resolution:
1 %
SpO
2
Accuracy:
70
~
100 %: ±2 % (non-motion)
70
~
100 %: ±3 % (motion)
0
~
69 % : Undefined
PR Measurement Range:
30
~
250 bpm
PR Resolution:
1 bpm
PR Accuracy:
±2 bpm (non-motion)
±3 bpm (motion)
Masimo SpO2
Measurement Range:
SpO
2
: 1
~
100 %
PR: 25
~
240 bpm
Perfusion: 0.02
~
20 %
SpO
2
Accuracy:
(non-motion)
1
Adult
Pediatric
70
~
100 %: ±2 %
0
~
69% : unspecified
Neonate
70
~
100 %: ± 3 %
0
~
69 % : unspecified
SpO
2
Accuracy:
(motion)
2,3
Adult
Pediatric
70
~
100 %: ±3 %
0
~
69% : unspecified
Neonate
70
~
100 %: ± 3 %
0
~
69 % : unspecified
PR Accuracy:
(non-motion)
1
Adult
Pediatric
Neonate
25
~
240 bpm: ±3 bpm
PR Accuracy:
(motion)
2,3
Adult
Pediatric
Neonate
25
~
240 bpm: ±5 bpm
SpO
2
Resolution:
1 %
PR Resolution:
1 bpm
Low Perfusion Performance
4
:
>0.02 % Pulse Amplitude
and % Transmission >5 %
SpO
2
: ±2 %
PR: ±3 bpm