SPO2 ERROR MESSAGES
PLETH Waveform may display messages as below:
PROMPTS
EXPLAINATION
Search Too Long
Search-time of SpO
2
is too long
Searching For Pulse. . .
Searching for pulse signal
Sensor Off
Sensor falls off or the finger fails to insert
into the finger-probe
SpO2 Com Error
SpO
2
board has communication error with
the mainboard
MASIMO INFORMATION
TRADEMARK AND LICENSING LABELS
MASIMO PATENTS
Please refer to Masimo
’s website for detailed information.
Masimo Patents: http:// www.masimo.com/patents.htm.
NO IMPLIED LICENSE
Possession or purchase of this device does not convey any express or implied license to
use the device with unauthorized sensors cables which would, alone, or in combination
with this device, fall within the scope of one or more of the patents relating to this device.
MEASUREMENTS
If the accuracy of any measurement does not seem reasonable, first check the patient
’ s
vital signs by alternate means and the check the MS board for proper functioning.
Incorrect measurements may be caused by:
[WARNINGS]
◼
Pulse rate measurement is based on the optical detection of a peripheral flow
pulse and therefore may not detect certain arrhythmias. So it should not be used
as a replacement or substitute for ECG based arrhythmia analysis.
◼
OMNI Express Patient Monitor should be considered an early warning device. As
a trend towards patient deoxygenation is indicated, blood samples should be
analyzed by a laboratory co-oximeter to completely understand the patient
’s
condition.
◼
Interfering Substances: Carboxyhemoglobin may erroneously increase readings.
The level of increase is approximately equal to the amount of carboxyhemoglobin
present. Dyes or any substance containing dyes that change usual arterial
pigmentation may cause erroneous readings.