ARRHYTHMIA ANALYSIS
The monitoring system supports the self-relearn function to accommodate itself to new
conditions such as different patients. The user can edit the arrhythmia type. For each type,
the system saves 8 items for arrhythmia and in total saves 104 items.
ARR ANALYSIS
Set arrhythmia analysis to be
ON
or
OFF
. The factory setting is
OFF
.
ARR SOURCE
Select between
lead I, lead II
and
Lead III
. The factory setting is
lead II
. The user can
switch the ECG lead if the current lead’s signal is weak.
PVC MONITOR
Set PVC monitor to ON or OFF. The factory setting is ON, and if the premature ventricular
contraction times exceed the PVC ALARM COUNT, the system will alarm.
PVC ALARM COUNT
The set range is from 1 to 10. The factory setting is 10.
ARR RELEARN
Self relearn to adjust to new conditions. For example, cardiographs can differ dramatically
between different patients.
ARR ALARM SETUP
Set all of the arrhythmia alarms to
ON
or
OFF
. The all factory settings are
ON
.
ARR REVIEW
Figure 12: Window for ARR Review
1.
Select:
To choose an arrhythmia item.
2.
Wave:
To review the selected arrhythmia item, including HR, ST, PR, SpO
2
, NIBP,
Temp, Resp, PVC and so on.
[WARNING]:
The Arrhythmia analysis program is intended to detect ventricular arrhythmias. It is not
designed to detect atrial or supraventricular arrhythmias. It may incorrectly identify the
presence or absence of an arrhythmia. Therefore, a physician must analyze the
arrhythmia information with other clinical findings.