Use
r’s
Manual of OMNI Express Patient Monitor
- 55
037.007.0001 Rev.2
ETCO2 SETUP
Touch the EtCO
2
Waveform or Parameter Area to open the menu for EtCO
2
Setup, which
is laid out as follows:
EtCO
2
Setup
Alarm Switch
EtCO
2
Alarm High
EtCO
2
Alarm Low
AWRR Alarm High
AWRR Alarm Low
Apnea Delay
EtCO
2
Unit
EtCO
2
Period
Zero Setup
Oxygen
BARO.
Advanced Setup
Gas Temp
Start Zeroing
Zero Gas Type
Balance Gas
Anesthetic
Fill In Waveform
Wave Scale
Sweep Speed
Sweep Color
Figure 22: Tree Diagram for EtCO
2
Setup Menu
ALARM SWITCH
Can be ON or OFF, and the factory setting is
ON.
When the alarm is ON, the alarm is
activated. When the alarm is
OFF
, the alarm indicator will not light up, the corresponding
alarm parameter will not flash and corresponding parameter area will display the
icon.
ETCO2 ALARM HIGH
The range is 20
~
100 mmHg, and the factory setting is 60 mmHg.
ETCO2 ALARM LOW
The range is 10
~
95 mmHg, and the factory setting is 15 mmHg.
AWRR ALARM HIGH