Use
r’s
Manual of OMNI Express Patient Monitor
- 39
037.007.0001 Rev.2
PATIENT TYPE
ADULT TYPE
This setting applies to the
adult
mode. Inflate the cuff to 180 mmHg (24 kPa). If the NIBP
value cannot be measured, then inflate the cuff for the second time, but note that the
maximum pressure cannot exceed 297 mmHg (40 kPa). The factory setting is ADULT
TYPE.
PEDIATRIC TYPE
This setting applies to the
pediatric
mode. Inflate the cuff to 170 mmHg (23 kPa). If the
NIBP value cannot be measured, then inflate the cuff for the second time, but note that the
maximum pressure cannot exceed 297 mmHg (40 kPa).
NEONATAL TYPE
This setting applies to the
NEONATAL
mode. Inflate the cuff to 100 mmHg (13 kPa). If the
NIBP value cannot be measured, then inflate the cuff for the second time, but note that
the maximum value cannot exceed 147 mmHg (20 kPa).
If this setup is done before the NIBP module is initiated, the settings will not be effective.
INFLATION TYPE
There are three choices. Manual, Auto and STAT.
MANUAL MODE
:
Press the
START/STOP
button to begin inflation. The information area of the display
should read
“Manual measuring… “. This shows that measurement is being taken.
Once the NIBP measurement is finished, NIBP parameter area will display the values and
the information area will read
“Manual measuring end!”, and the measurement process
will have finished.
If the NIBP value cannot be measured, the NIBP parameter area will display any error
messages and automatically attempt to take the measurement again up to three times, If
the value still cannot be measured, the information area will read
“RETRY OVER
!
” and no
more measurements will be taken.
During the measurement, press the
START/STOP
button again to stop the NIBP
measurement process. The information area will read "Stop manual measuring".
AUTOMATICAL MODE
NIBP parameter area will display the countdown of "Auto measuring..." (Time Interval).
Under this setting, the machine will continue re-measure NIBP at every time interval until
the mode is changed.
If you start a measurement manually, the monitor will then continue automatically repeat
NIBP measurements at the set time interval.
If the NIBP measurement is finished, the NIBP parameter area will display the values and
the information area will give a note of "Auto measuring end". The monitor will then
continue automatically measuring until the mode is changed.
If the NIBP value cannot be measured, the NIBP parameter area will display some error
messages and will attempt to take the measurement again up to three times. If the value
still cannot be measured, the information area will read
“RETRY OVER
!
” And
automatically continue to take measurements at every time interval, until the mode is
changed.
If the
START/STOP
button is pressed during any point in the countdown period, the