8 – 9
Transpector SPS Operating Manual
8.4 If You Cannot Resolve Your Problem
Contact INFICON. (Refer to
.)
8.5 Event Log Files
FabGuard Explorer software records Errors and Events with time of occurrence for
each day of operation, providing a chronology of operations that can be useful in
diagnosing problems. See FabGuard Explorer Operating Manual (PN 074-528-P1)
for additional information.
High noise
level
System grounding
Verify that vacuum system
is grounded
Electronics failure
Return to INFICON for
repair
Transpector SPS
electronics module not
mounted properly on
sensor
Ensure Transpector SPS
electronics module is fully
seated on sensor
EM defective
Replace EM assembly or
sensor
Scan speed too fast for
gain setting
Increase dwell time
Table 8-1 Symptom—Cause—Remedy Chart (continued)
SYMPTOM
CAUSE
REMEDY