10 – 19
Transpector SPS Operating Manual
10.6.2.3 Stop Parameters
Defines how and when the data acquisition stops.
10.6.2.3.1 Stop Mode Type
Defines when the data acquisition stops. (See
.)
Figure 10-32 Stop mode SPS
Time (s)
stops the acquisition after a set period of time.
Maximum Signal
stops the run after the sensor sees a defined maximum
signal. The signal is set by the Data Threshold. (See
.)
Pressure Above
stops the run immediately after an increase in the internal
RGA total pressure above the user-defined
Data Threshold
.
Pressure Below
stops the run immediately after a decrease in the internal
RGA total pressure below the user-defined
Data Threshold
.
Analog 1
Above
stops the run after an increase in voltage above the
user-defined Data Threshold.
Analog 1
Below
stops the run after a decrease in voltage above the
user-defined Data Threshold.
Number of Points
stops the run after collecting a user-defined number of
scans.
10.6.2.3.2 Data Threshold
Defines the threshold that the
Stop Mode Type
utilizes to determine when the run
will stop. The
Data Threshold
is set in units corresponding to the
Start Mode Type
selected.
Maximum Signal
is in units of current,
Pressure Above/Below
is in
units of pressure, and
Analog 1 Above/Below
is in units of voltage. (See
.)
Figure 10-33 Data threshold (stop mode)