10 – 12
Transpector SPS Operating Manual
10.6.1.2.1 Existing Masses
Editing Masses
If a mass has already been added but the mass has the wrong identity for the
process, click
Edit
to open the
Mass Properties
window and change the identity.
For example, a process may require monitoring of carbon monoxide. Since carbon
monoxide has the same mass as nitrogen (28 amu), the identity of mass 28 must
be changed from nitrogen to carbon monoxide in the
Identity
list.
Dwell time can be changed in the
Dwell
list.
In the
Relay
list, thresholds can be set up to open or close relays based on the
signal of the mass bin. (See
.)
Figure 10-17 Mass properties
CAUTION
Editing a special bin, such as Ionizer State, will display an
error message. (See
.)