1 – 7
Transpector SPS Operating Manual
1.6.2 Weight
The approximate weight of Transpector SPS (without the foreline pump and the
Controller) is 15.5 kg (34 lb.).
NOTE:
This weight does not include connecting cables.
1.6.3 Ventilation Requirements
CAUTION
For adequate ventilation, maintain at least 25.4 mm (1 in.)
clearance around the Transpector electronics module,
the Transpector SPS heater and the pumping system.
If enclosing Transpector SPS, the enclosure must be large enough or ventilated to
provide adequate cooling airflow to the fan on the Transpector SPS pumping
system and the fan in the Transpector SPS controller.
1.7 Electrical Power Requirements
The Transpector SPS system components that require AC power input are:
Controller . . . . . . . . . . . . . . . . . . . . . Universal input, any voltage in the range
specified in
is acceptable
Personal computer . . . . . . . . . . . . . . Universal input, any voltage in the range
specified in
is acceptable
1.7.1 Required Supply Voltage
Except for the AC power input, the Controller supplies all required voltages.
1.7.1.1 Acceptable Supply Voltage Range
The AC power input must be 100, 120, or 230 V
~
,
±10%
NOTE:
If the input power is less than 100 V
~
, the time required for the heater to
reach its nominal temperature will be increased.
1.7.1.2 Required Frequency
50 or 60 Hz, ±5%
1.7.1.3 Power Rating
Controller—500 VA