ISO 14117:2019
Clause 4.9 - Test
requirements for the
frequency range of
385 MHz ≤ ƒ ≤ 3000
MHz
Per standard
Functions as it
did before the
test without
further
adjustment
after application
of the test
signal per
standard
Seek the advice of your
physician or other qualified
health provider regarding
Transmitting Devices
and
Cellular and Mobile Phones
•
Exercise caution in
the vicinity of
equipment that
generates strong
radio
-frequency
fields.
•
Do not enter an
area with posted
warnings advising
pacemaker patients
(or patients with
other types of
implantable
devices) not to
approach.
•
Interference may
occur in the vicinity
of equipment
marked with the
following symbol:
ISO 14117:2019
Clause 5 - Testing
above frequency of
3000 MHz
Standard does
not require
testing of
devices above
3
GHz.
Electromagnetic
fields > 3 GHz
are not
expected to
interfere with
device operation
because of the
increased
devi
ce
protection
afforded by the
attenuation of
the enclosure
and body tissue
at microwave
N/A
Avoid direct exposure to the
main lobe of high
-
power radar
and microwave
communication beams.