You should avoid getting too close to equipment or devices
that
contain strong magnets (e.g.,
stereo speakers)
or leaning over an
open automobile engine compartment, as the alternator generates
a strong electromagnetic field
. Your OPTIMIZER Smart Mini
IPG
contains a magnetic switch that, when
expos
ed
to a strong
magnet for 3-
5 seconds, will disable CCM therapy delivery
. If this
occurs by accident, your doctor
may require you to come to their
office to restore CCM therapy delivery
. Since your OPTIMIZER
Smart Mini
IPG
is not a life-
sustaining device, you are unlikely to
be placed at risk by such an event.
Always
seek
medical advice before entering an area posted with a
warning for pacemaker patients (or other medical implantable
devices) or where there is industrial machinery or radio
transmitters, including ham radios and mobile radios.
Always
inform your doctor that you have an implanted
OPTIMIZER Smart Mini
IPG
before
you undergo the following
procedures:
•
S
urgery where electrocautery is going to be used
•
A
procedure involving Radiofrequency (
RF
)
Ablation
•
Medical
Diathermy
•
Cardi
oversion
•
Therapeutic R
adiation
•
Therapeutic U
ltrasound
•
Lithotripsy
•
N
uclear Magnetic Resonance (NMR)
•
M
agnetic Resonance Imaging (MRI)
Warning:
DO NOT
undergo an MRI procedure when implanted
with an OPTIMIZER Smart Mini IPG
.
Caution:
Your OPTIMIZER Smart Mini
IPG should either be
deactivated or closely monitored prior to and during any medical
treatment in which electrical current is passed through the body.
Caution:
Your OPTIMIZER Smart Mini
IPG should not be directly
exposed to therapeutic ultrasound or therapeutic radiation
. This
type of exposure may damage the device
that
may not be
immediately detectable.