75
1693 RLC Digibridge
Operation
digits (0 to 9) which convey information about the
selection within the category. For example, the
ASCII sequence D2 means “in DISPLAY category,
select VALUE.” The command is like pressing the
[DISPLAY] key to select VALUE.
Floating Point Entries. These entries (also shown in
the table) are ASCII character sequences of arbitrary
length, always terminated with a semicolon (;). The
first character is a letter which designates the category
of the entry. The following ASCII characters - before
the semicolon - define a floating-point number includ-
ing at least one decimal digit (0 to 9) and optional
characters (+ - . e E). Any space character is ignored.
The character e or E is recognized as “exponent” in
E-format notation. For example, any of the following
three entries will set the test frequency to 100 Hz:
F0.l; F1e-1; F100.0E-3;
NOTE: The 2-byte command P2 is recommended as
the first command in a remote programming sequence
because it resets all selections to their power-up (de-
fault) conditions.
3.11.8 Legacy Digibridge IEEE-488
Commands
The legacy IEEE-488.1 Digibridge commands are
supported by the IEEE-488 interface board for com-
patibility with software and drivers written for the
original Digibridge IEEE-488.1 interface boards.
Category
Description
Type
Code
Display
Bin #
2 byte
D0
Delta %
2 byte
D1
Value
2 byte
D2
Delta RLC
2 byte
D3
Ratio: nominal value /
RLC value
2 byte
D4
Ratio: RLC value /
nominal value
2 byte
D5
Bin on ratio: nominal
value / RLC value
2 byte
D6
Bin on ratio: RLC value
/ nominal value
2 byte
D7
Measurement
Rate
Fast
2 byte
S0
Medium
2 byte
S1
Slow
2 byte
S2
Category
Description
Type
Code
Measurement
Voltage
Constant, off
2 byte
Y0
Constant, on
2 byte
Y1
Value (in volts) ═ v
Floating
point
Vv;
Function
Measure
2 byte
P0
Enter
2 byte
P1
Reset
Turn-On Conditions
2 byte
P2
Diagnostics
2 byte
P3
Measurement
Mode
Triggered
2 byte
T0
Continuous
2 byte
T1
Triggered median value
2 byte
T2
Continuous median
value
2 byte
T3
Parameter
Inductance (L/Q)
2 byte
M0
Capacitance (C/D)
2 byte
M1
Capacitance (C/R)
2 byte
M2
Resistance (R/Q)
2 byte
M3
Inductance L/Q) ppm
2 byte
M4
Capacitance (C/D) ppm
2 byte
M5
Capacitance (C/R) ppm
2 byte
M6
Resistance (R/Q) ppm
2 byte
M7
Automatic selection of
R/Q, C/D, or L/Q
2 byte
M8
Resistance (R/X)
2 byte
K2
Conductance (G/B)
2 byte
K3
Impedance (Z/ANG)
2 byte
K4
Admittance (Y/ANG)
2 byte
K5
Impedance (Z/ANG)
ppm
2 byte
K6
Admittance (Y/ANG)
ppm
2 byte
K7
Equivalent
Circuit
Parallel
2 byte
C0
Series
2 byte
C1
Range Control
Hold range
2 byte
R0
Hold range 1
2 byte
R1
Hold range 2
2 byte
R2
Hold range 3
2 byte
R3
Hold range 4
2 byte
R4
Autorange
2 byte
R5
Frequency
Value (in kHz) ═ f
Floating
point
Ff;
Zero Calibration
Disable
2 byte
Z0
Enable open
2 byte
Z1
Enable short
2 byte
Z2
Output Format
Normal
(not
compacted)
2 byte
x0 *
Compacted binary
format
2 byte
x1 *
Summary of Contents for RLC Digibridge 1693
Page 12: ...x Table of Contents 1693 RLC Digibridge This page is intentionally left blank...
Page 24: ...xxii 1693 RLC Digibridge Abbreviated Specifications This page is intentionally left blank...
Page 130: ...106 1693 RLC Digibridge Theory Figure 4 1 Block diagram of the 1693 RLC Digibridge...
Page 171: ......