91
1693 RLC Digibridge
Operation
3.11.11 Example Programming
Zero-Open
ibwrt “SYST:RES”
ibwrt “CONF:PAR CD”
ibwrt “CONF:OUTP RLC”
ibwrt “CONF:CIRC PAR”
ibwrt “CONF:AVER 10”
ibwrt “SYST:CAL OPEN”
ibwrt “MEAS”
ibrd 100
Zero-Short
ibwrt “SYST:RES”
ibwrt “CONF:PAR RQ”
ibwrt “CONF:OUTP RLC”
ibwrt “CONF:CIRC SER”
ibwrt “CONF:AVER 10”
ibwrt “SYST:CAL SHORT”
ibwrt “MEAS”
ibrd 100
3.11.12 Talk-Only Use, for Data
Output
This pertains to a relatively simple system, with the
Digibridge outputting data to one or more “listen-on-
ly” (IEEE-488 compatible) devices such as a printer.
1. Set the TALK switch to TALK ONLY.
2. Program the Digibridge to send out results
automatically after each measurement. (Refer
to paragraph 3.11.3.)
The “special” commands for this purpose can be
executed only in ENTER function, as follows.
Bin numbers only
[1]= [SHIFT][SPECIAL][2]
QDR results only
[2]= [SHIFT][SPECIAL][2]
QDR and bin numbers
[3]= [SHIFT][SPECIAL][2]
RLC results only
[4]= [SHIFT][SPECIAL][2]
RLS and bin numbers
[5]= [SHIFT][SPECIAL][2]
RLC and QDR results
[6]= [SHIFT][SPECIAL][2]
RLS, QDR and bin numbers
= full data
[7]= [SHIFT][SPECIAL][2]
No data on PASS, Full data
on FAIL
[8]= [SHIFT][SPECIAL][2]
BIN No. on PASS, full data
on FAIL
[9]= [SHIFT][SPECIAL][2]
For no data output via IEEE-
488 bus
[0]= [SHIFT][SPECIAL][2]
Operate the Digibridge in the usual way (manually).
The system may constrain operation in some way. For
example, a slow printer will limit the measurement
rate because it requires a certain time to print one
value before it can accept the next.
Refer to Table 3-20 for examples of the interface
message code for data transfer. This table shows
typical codes, but does not repeat the entire ASCII
code. (There is a more complete table, in paragraph
Table 2-1 on page 11.)
NOTE: For a description of the compacted binary data
output format (which saves time), refer to paragraph
3.11.14.
Character
ASCII
Decimal
Equivalent
DI08
DI07
DI06
DI05
DI04
DI03
DI02
DI01
A
65
0
1
0
0
0
0
0
1
B
66
0
1
0
0
0
0
1
0
C
67
0
1
0
0
0
0
1
1
*** etc ***
1
49
0
0
1
1
0
0
0
1
2
50
0
0
1
1
0
0
1
0
3
51
0
0
1
1
0
0
1
1
*** etc ***
Table 3-20: Abbreviated interface message code for “TALK-ONLY” data transfer
Summary of Contents for RLC Digibridge 1693