xix
1693 RLC Digibridge
Abbreviated Specifications
Binning:
Pass bins:
13 pass bins for RLCGZY
Fail bins:
2 fail bins, RLCGZYΘ
Sorting capabilities:
Bin number, Delta RLC, Delta %, Value
DUT connections:
The 1693 has four BNC connections to make
four-terminal (Kelvin) connections to the device
under test. These connections are shielded and the
instrument ground has a guard for three-terminal
measurements. 1689-9602 1-meter BNC Cable
Set is provided. Various test fixtures and other
accessories are available.
Front of
1693
Description 1689-9602
BNC Cable
1689-9600 Remote
Test Fixture*
IL
Drive -
IL/Black Band
I+
PL
Potential -
PL/Black and
White Band
P+
PH
Pot
PH/Red and
White Band
P-
IH
Drive +
IH/Red Band
I-
*High and Low Reversed for correct operation
Charged capacitor protection:
The instrument is protected from damage due to the
connection of charged capacitors with up to 1 joule
of stored energy:
Stored Energy = ½CV
2
Joules
Keyboard lock:
The keyboard can be locked to prevent inadvertent
changes to test conditions and sorting routines.
Parameter storage:
All keyboard settings and programmed nominal
values and bin limits can be stored in non-volatile
memory for automatic reentry upon power up .
Self-Check diagnostics:
Self-tests are performed during power-up to verify
proper operation and validity of calibration. Coded
error signals on the display notify the operator if a
problem is encountered.
Special functions:
Several special features may be selected including:
• Direct Range Setting
• Range Extension (RATIO mode)
• Integration Time
• Blanking of lesser digits
• Signal reversal for hum removal under adverse
conditions
• A Median value routine
• Automatic parameter selection
• Quick acquisition routine
• Choice of data output on IEEE bus
• Interface Option: IEEE-488 Bus (1689-9640)
Zeroing:
Open and short circuit compensation
Power:
90-250 Vac
50-60 Hz
60 W max
Environmental conditions:
Operating conditions:
0° to +50°C, <85% RH
Storage conditions:
-45°C to +75°C
Dimensions:
43.8 cm W x 14.3 cm H x 38.5 cm D (17.3” x 5.6”
x 15.2”)
Weight:
6.4 kg (14 lbs)
Supplied accessories:
Power cable
1689-9602 bnc-to-bnc extender cable, 1-meter
Instruction manual
Calibration certificate traceable to SI
Optional accessories:
1700-03
Kelvin Test Leads
7000-04
Alligator Clip Leads, 1 Meter
7000-05
Chip Component Tweezers
1689-9602-2 bnc-bnc Extender Cable, 2-Meter
1689-9600 Remote Test Fixture
1689-9640 IEEE Digibridge Interface
1689-9604 Digibridge Calibration Kit
1689-9611 Rackmount Kit
1657-9600 Banana-Plug Extender Cable
1658-2450 Bias cable
Calibration certificate accredited to ISO17025
Summary of Contents for RLC Digibridge 1693
Page 12: ...x Table of Contents 1693 RLC Digibridge This page is intentionally left blank...
Page 24: ...xxii 1693 RLC Digibridge Abbreviated Specifications This page is intentionally left blank...
Page 130: ...106 1693 RLC Digibridge Theory Figure 4 1 Block diagram of the 1693 RLC Digibridge...
Page 171: ......