1693 RLC Digibridge
Abbreviated Specifications
Condensed operating instructions
1. General Information
The 1693 RLC Digibridge is a microprocessor-
controlled, automatic RLC meter.
Measurement parameters:
R/Q, L/Q, C/D (series
or parallel), R/X (series), G/B (parallel), Z/Angle or
Y/Angle
Ranges:
Parameter
Direct Reading Range
Ratio and DQ in PPM
R and [Z]
0.00001 Ω to 99999 kΩ
0.00010 Ω to 9999.9
GΩ
L
0.00001 mH to 99999 H
0.00010 nH to 9999.9
MH
C
0.00001 pF to 99999 µF
0.00010 aF to 9999.9 F
Test frequencies:
Over 500 selectable test
frequencies ranging from 12 Hz to 200 kHz
Applied voltage:
5 mV to 1.275 V (programmable
in 5 mV steps)
Measurement speed:
Up to 19 measurements per
second
Bias:
Internal:
2.0 Vdc;
External:
60 Vdc max
Accuracy: Basic RLCGZY:
±0.02%
Sense +
Sense -
Drive +
Drive -
Match spacing of
DUT terminals
DUT
+
-
Sense +
Sense -
Drive +
Drive -
Open
Connection
Short
Connection
2. Starting-up
a. If the Digibridge includes an optional IEEE-488 interface,
set [
TALK]
switch (rear panel) to
TALK ONLY
(unless
instructions are to be received through the IEEE-488 bus).
b. Make sure the [
EXTERNAL BIAS]
is in the
OFF
posi-
tion.
c. Press the [
POWER]
button.
The unit runs a brief self-check on start-up.
d. Wait until keyboard lights indicate
MEASURE
,
VALUE
,
SLOW
,
TRIGGERED
,
SERIES
.
If a fault is detected, measurements are blocked and an er-
ror code remains displayed. See manual, paragraph 3.13.
If keyboard lights remain dark, keyboard is locked. To
unlock it, see manual paragraph 3.9.
To switch power off, press the [
POWER]
button.
3. Obtaining optimal accuracy by zeroing
Before making measurements, zero the Digibridge as fol-
lows: (Note: For best accuracy, repeat this procedure every
day, after each change of frequency, and after any change
of test-fixture adaptors.)
a. Attach extender cables or whichever test fixture/adapter is
appropriate.
b. Create an open circuit by having the spacing b
and - cable ends match the spacing of the DUT terminals.
See the figure.
If the spacing of DUT terminals is unknown, make sure
that the + and – ends of the cable are at least 0.75” (19
mm) apart.
c. Confirm that
MEASURE
keyboard light is lit.
d. Press [
MEASURE MODE]
key to select
TRIGGERED
mode
e. Press the following keys:
[1] [6] [9] [3] [=] [SHIFT]
[OPEN]
.
f. Press
[START]
and wait for the green
GO
light.
g. Create a short circuit across test terminals by connecting all
cable ends together as shown in the figure.
h. Press the following keys:
[1] [6] [9] [3] [=] [SHIFT]
[SHORT]
.
i. Press
[START]
and wait for the green
GO
light.
At this point, the unit should be ready for taking measure-
ments. If something goes wrong during the zeroing process,
the procedure should be repeated.