127
1693 RLC Digibridge
Service and Maintenance
5.7.3 Internal Fuse Replacement
If at power-up, there is a failure display of
555 5:XXXX (see Self Check paragraph 3.13.1 and
Trouble Analysis paragraph 3.7.1) replacement of the
capacitor discharge fuse may be necessary.
WARNING
Take appropriate safety precautions
when servicing the unit.
Internal Fuse
Figure 5-11: Location of internal fuse
1. Remove the top and bottom covers.
2. This fuse is located in a socket on the main
board and behind the power switch. Unplug
the fuse (using needle nose pliers) and
replace.
3. Install top and bottom covers.