93
1693 RLC Digibridge
Operation
Character
Sequence
Purpose
Allowed
Characters
Meaning
1
Status
(space)
Normal Operation
O
Overrange held (reduced accuracy)*
I
Invalid measurement due to signal overload
2
Format
(space)
3
Parameter
Q
Quality factor
D
Dissipation factor
R
Resistance
X
Reactance
B
Susceptance
A
Angle
4
Format
(space)
5..7
Units
(2 space)Ω
Ohms
(space)kΩ
Kiloohms
(2 space)S
Siemens
(space)mS
Millisiemens
(space)µS
Microsiemens
deg
Degrees
ppm
Part per million (for D, Q or Angle)
(3 space)kO
Dimensionless (for Q or D)
7
Format
(space)
8
Sign
(space)
Positive Value
-
Negative Value
9..15
Number
0123456789.
(space)**
Measured number, right justified in format field, like the QDR
display except the zero before the decimal point is explicitly
provided and this number can be as long as six characters. *
16
(CR)
The standard “carriage return” and “line feed” characters;
end of string.
17
(LF)
Table 3-22: QDR-Value data output format
* If measurement is overrange (display goes blank), 999999 is output in this field.
** Two leading zeros will precede the decimal point on values less than 1. Two spaces will precede the value on values greater
than 1.
Character
Sequence
Purpose
Allowed
Characters
Meaning
1
Pass/fail
(space)
Go (bins 01 to 13)
F
No-Go (bins 00 or 14)
2
Format
(space)
3
Label
B
The word “BIN”
4
I
5
N
6
Format
(space)
7,8
Bin Number
123456789
Bin number assignment, 00 to 14
9
(CR)
The standard “carriage return” and “line feed” characters;
end of string.
10
(LF)
Table 3-23: BIN-Number data output format