1693 RLC Digibridge
Operation
3.11.9 IEEE 488.2/SCPI Digibridge
Command Summary
The IEEE-488.2 / SCPI commands are used to com-
municate with the Digibridge via the IEEE-488 or
RS232 serial interfaces replacing and/or supplement-
ing the legacy commands.
Command
Parameter(s) Description
Legacy
CONFigure:
PARameter
Selects the measured parameter
LQ
Inductance
(L/Q)
M0
CD
Capacitance
(C/D)
M1
CR
Capacitance
(C/R)
M2
RQ
Resistance
(R/Q)
M3
LQPPm
Inductance
(L/Q) ppm
M4
CDPPm
Capacitance
(C/D) ppm
M5
CRPPm
Capacitance
(C/R) ppm
M6
RQPPm
Resistance
(R/Q) ppm
M7
AUTo
Automatic
selection of
R/Q, C/D, or
L/Q
M8
RX
Resistance
(R/X)
K2
GB
Conductance
(G/B)
K3
ZANG
Impedance (Z/
ANG)
K4
YANG
Admittance (Y/
ANG)
K5
ZAPPm
Impedance (Z/
ANG) ppm
K6
YAPPm
Admittance (Y/
ANG) ppm
K7
CIRCuit
Selects the measured equivalent circuit
PARallel
Parallel
C0
SERies
Series
C1
APERture
Measure rate
FAST
Fast
S0
MEDium
Medium
S1
SLOW
Slow
S2
Command
Parameter(s) Description
Legacy
MODe
Sets the measure mode
TRIGgered
Triggered
T0
CONTinuous
Continuous
T1
MTRiggered
Triggered
median value
T2
MCONtinuous
Continuous
median value
T3
DISPlay
Results to display on the front panel
BIN
Bin #
D0
DELTa
Delta %
D1
VALue
Value
D2
DRLC
Delta RLC
D3
:RATio
NOMValue
Ratio: nominal
value / RLC
value
D4
VALNominal
Ratio: RLC
value / nominal
value
D5
:BINRatio
NOMValue
Bin on ratio:
nominal value /
RLC value
D6
VALNominal
Bin on ratio:
RLC value /
nominal value
D7
FUNCtion
Sets function mode to measure or entry setting
MEASure
Measure
P0
ENTer
Enter mode
for setting
configuration
P1
RANGe
Sets the measurement range
HOLD
Hold range
R0
1
Hold range 1
R1
2
Hold range 2
R2
3
Hold range 3
R3
4
Hold range 4
R4
AUTo
Autorange
R5
DELay
Sets the trigger delay time
<numeric>
Real value in
ms
J<numeric>;
ITFactor
Sets the integration-time multiplying factor
<numeric>
Real value
multiplies
Medium and
Fast
I<numeric>;
integration times
AVERage
Sets the number of measurements to average
<numeric>
# of
measurements
A<numeric>;
FREQuency
Sets the frequency used for measurement
<numeric>
Real value in
kHz
F<numeric>;
VOLTage:
Sets the measurement signal voltage