Technical data and symbols
46
14
Technical data and symbols
Technical data
The blood pressure monitor complies with the EMC directives.
The blood pressure cuff and air hose are made of a non-conductive material. This makes the device defibrillator
safe.
Specification
Value
Unit
Measuring method
Oscillometric
Blood pressure
measurement range
Systolic
60 to 290
Diastolic
30 to 195
mmHg
Pulse measurement range
30 to 240
1/min
Pressure Accuracy
±2% or ±3 mmHg, whichever is greater
Memory (without PWA)
350
Measurements
Memory (with PWA)
15
Measurements
Power supply
4-6 VDC (4x NiMH or LR6, AA)
Dimensions (L x W x H)
151 X 108 X 57
Mm
Weight (without batteries)
344
G
Material (housing)
ABS (acrylonitrile-butadiene-styrene)
Material (cuff)
Polyester
IP protection class
20
Operating temperature
+5 to +40
°C