Manufacturer’s EMC guidelines
52
Elektromagnetic immunity
The Tel-O-Graph
®
GSM is intended to be operated in the electromagnetic environment specified below. Only use
the Tel-O-Graph
®
GSM in such an environment.
Measurement of
interference
emissions
Test level
Compliance level
Electromagnetic environment
–
guidelines
Electrostatic
discharge (ESD) in
accordance with
IEC 61000-4-2
+/- 8kV
contact discharge
+/- 15kV
air discharge
+/- 8kV
contact discharge
+/- 15kV
air discharge
Floors should consist of wood or
cement or ceramic tiles. If the floor
consists of synthetic materials, relative
humidity must be at least 30%.
RF radiated
disturbances in
accordance with
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
Do not use wearable or portable radio
equipment closer to the Tel-O-
Graph
®
GSM or its cables than the
recommended safe distance.
The field strength of stationary radio
wireless transmitters should be lower
than the compliance level at all
frequencies, as shown by an on-site
examination. Interference is possible in
the vicinity of devices bearing the
graphic symbol
“
non-ionising radiation
”
.
IEC 61000-4-4
Not applicable
IEC 61000-4-5
Not applicable
IEC 61000-4-6
Not applicable
Summary of Contents for Tel-O-Graph GSM
Page 2: ...030500001 Ver 1 0 30 07 2021 ...
Page 58: ......
Page 114: ......
Page 170: ......
Page 226: ......
Page 282: ......
Page 338: ......
Page 339: ......
Page 340: ...030500001 Ver 1 0 30 07 2021 ...