Example of a formula
Ch5/600*(Ch1-Ch2)
Press the
key once to enter a
Ch
(= channel), resp. press twice to enter a
5
. You can only enter numbers with the remaining number keys; special char-
acters with
.
Confirm the input with
. The measuring system does not check the entered
formula for plausibility.
Example of a
consumption
measurement in [l/min]
Some measurement channels are required for this example. They are printed
in
bold
letters:
•
Channel 4
: Measurement of volume V1 in liters (l)
•
Channel 5
: Measurement of volume V2 in liters (l)
•
Channel 11: Calculation
Ch7 - Ch8 = dV1
in liters (l)
•
Channel 12: Calculation
dCh11/dt = Q1
in liters per second (l/s)
•
Channel 13: Calculation
Ch12 * 60 = Q2
in liters per minute (l/min)
Values from special channels can only be used if the ordinal number of the used
channel is lower.
Possible formula on channel 14:
Ch12+Ch1
, impossible formula on Ch14:
Ch15+Ch1
.