Firmware Version 1.2
MultiSystem 4070
31
Operation
DEU
Available measurands
The instrument is able to process a number of different measurands including
pressure, volume flow rate, temperature and rotational speed. Make sure you
select the measurand and unit corresponding to the sensor.
Index Measurand
If several channels are programmed with the same measurand, these will be
automatically indexed consecutively. The automatic indexing can be disabled
in the Device menu to allow manual assignment of index numbers.
Channel name
You can assign an individual name to each channel.
Signal types
Choose between
0/20 mA
,
4/20 mA
,
0/10 V
,
± 10 V
,
0.5/4.5 V
,
1/5 V
and
2/10 V
.
On the frequency channels (Ch4-Ch5) you can also choose between
FRQ
,
± FRQ, CNT, ± CNT
and
± CNT (4Q)
.
Measuring range
Enter the beginning and end of the measuring range. Confirm these two en-
tries with
.
Zero point
Press
and START
to execute the automatic zero point equalization. A
possible zero point deviation will be compensated by the software.
Linearisation
If a calibration table is available for the connected sensor, you can enter it here
after selecting
YES
at the
Linearization
dialog entry.
Please observe the additional information in chapter Kapitel
LOAD
Press
to load sensor parameters from the sensor data base.
SAVE
Press
to save the current sensor parameters in the sensor database.