![ICS HYDROTECHNIK MultiSystem 4070 Operating Instructions Manual Download Page 75](http://html1.mh-extra.com/html/ics/hydrotechnik-multisystem-4070/hydrotechnik-multisystem-4070_operating-instructions-manual_615068075.webp)
Firmware Version 1.2
MultiSystem 4070
75
Operating software
DEU
POWER
uses the formula
Ch1 x Ch7 / 600
to calculate the hydraulic power
The pressure p in bar is measured on channel 1 and the volume flow rate Q in
l/min is measured on channel 7
FORMULA
Input of an individual formula
See
CAN
Observe the information in chapter
Viscosity
For viscosity calculation
See
Viscosity-compensated volume flow rate measurement
.
Set up the MultiXtend
1
Select the
Calculation
dialog entry:
2
Select the
MultiXtend
entry:
3
Select the
Signal type
dialog entry:
4
Select the MultiXtend model:
5
For the multi-channel MultiXtend model, select the
Port
dialog
entry:
6
Select the port corresponding to the desired channel:
7
Confirm changes and exit dialog:
OK
Confirms input/saves change.