![ICS HYDROTECHNIK MultiSystem 4070 Operating Instructions Manual Download Page 74](http://html1.mh-extra.com/html/ics/hydrotechnik-multisystem-4070/hydrotechnik-multisystem-4070_operating-instructions-manual_615068074.webp)
Firmware Version 1.2
MultiSystem 4070
74
Operating software
DEU
Channel name
You can assign an arbitrary name here
Align. Diff
This function automatically determines the measured value difference be-
tween the selected channels and uses it as offset.
Formula
Enter the desired formula here (only displayed if
Calculation
is set to
FOR-
MULA
, see
Possible assignments of the special channels
Possible assignments of the special channels
UNDEF
Channel is not in use
Ch1-Ch2
forms the difference of the measured values from channel 1 and channel 2
(Delta-x)
Here, both channels must be assigned with the same measurand and unit; the
resulting measurand and unit are determined automatically
the same applies to the assignments
Ch3-Ch4
,
Ch5-Ch6
and
Ch7-Ch8
dCh1/dt
forms the first derivative of the measured values from channel 1
analogously, the derivation of the channels Ch7 (
dCh7/dt
) and Ch11 (
dCh11/
dt
) is also possible
LOAD
Loads sensor parameters from the database.
SAVE
Stores the current sensor parameters in the database.
OK
Confirms input/saves change.