FCU 8000 series
FCU Display
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 85/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
If the determined particle count lies above the specified display range (see
page 109), >25 or >15 is shown in all particle size ranges as the particle
count.
5
One can select between differential and cumulative displays of numbers of
particles by using the keys
. The selected presentation is indicated
in the display by a "D" (differential) or "C" (cumulative) behind the display of
the expired measurement time.
Differential representation means particle count in the particle size ranges:
FCU 81xx:
2…5µm, 5…15µm, 15…25µm, 25…50µm, 50…100µm and >100µm
FCU 82xx:
4…6µm
(c)
, 6…14µm
(c)
, 14…21µm
(c)
, 21…38µm
(c)
, 38…70µm
(c)
and >70µm
(c)
Cumulative representation means particle count in the particle size ranges:
FCU 81xx:
>2µm, >5µm, >15µm, >25µm, >50µm and >100µm
FCU 82xx:
>4µm
(c)
, >6µm
(c)
, >14µm
(c)
, >21µm
(c)
, >38µm
(c)
and >70µm
(c)
If the determined particle count lies above the specified display range (see
page 109), ">>>>>>>“ is shown in all particle size ranges as the particle
count.