FCU 8000 series
FCU menus
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 75/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
Setting the pump start delay
When conducting measurements with the aid of an external pump or an external
filtration unit (e.g. OF5C), this function enables the user to operate the pump for a
limited period of time within which a flow has to start at the FCU.
When the pump start delay has elapsed and the FCU does not detect any flow,
measurement is stopped and the unit connected via a relay is switched off so as to
prevent damage caused by dry running of the pump. This function is in effect in all
operating modes except M2.
Measurement resumes as soon as enough flow is available. The stopping time
ranges from 1 to 200 seconds (practical recommendation: 60 seconds).
The 0 sec. setting causes the averaging function to be deactivated. This means that
the FCU waits as long as you wish for flow
setup
SETUP
Pump start delay
ok
start
Pump start delay:
xx
s
or
Pump start delay:
xx
s
ok
start
ok
start
Press this key to confirm the
operating mode selected and to
initiate measurement.
stop
Abort and exit.