FCU in BSU operation (only for FCU 8110 / 8210)
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 34/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
Perform basic settings for bottle sampling operation
The following settings must be made in the
POWERUP
menu under point "M6:
Bottle Sampler" (see also page 80)
BottleSampler draining
When the flushing procedure is completed in
mode M5, or the measurements in mode M6,
the drain valve will open for the configured time
so that the oil can completely drain from the
test container (see 44). (Recommended:
30 seconds)
Compressor
At this point you must pre-set the time required
for a constant pressure to be achieved in the
pressure chamber, after switching on the
compressed air supply via a compressor.
(We recommend 30 seconds using the
CompressedAir Unit CAU 8110)
Compressed air
At this point you must pre-set the time required
for a constant pressure to be achieved in the
pressure chamber, after switching on the
compressed air supply from the ring main.
(Recommended: 5 seconds)
The following settings must be made in the
SETUP
menu (see also page 72)
BS compressed air supply
At this point the type of compressed air supply
to the BSU must be specified.
Options: "Compressor" or "Compressed air"
The relevant pressure build-up time, which is
specified in the POWERUP menu, is used as a
basis for the modes of operation of bottle
sampling.
Additional test
At this point you have the option whether to
make an additional individual test in mode M6,
once the pre-set number of individual tests
have been completed.
Options: "Yes" or "No"