FCU 8000 series
FCU in BSU operation (only for FCU 8110 / 8210)
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 37/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
Once each individual measurement has been completed, the relevant result is
printed out when online printing is activated (see page 53).
The resultant quantity of particles is projected for 100 ml.
If "Yes" was selected in the Setup-Menu under additional measurement, you will be
asked whether an additional single measurement (measurement volumes as for
completed individual measurements) should be carried out, once the last individual
measurement has ended.
Display:
mode
M6 +1Measurement -> OK
finish ->STOP
After the last measurement has been made or after the button
stop
has been
pressed, the pressure chamber is depressurized and the test container drained.
On the transparent drain hose, check that the test container drains empty and no
medium gets trapped. With high-viscosity media, this process can take some time. If
necessary, correct the time for the BS draining in the POWERUP menu upward.
The single measurements which are to be used to produce an average overall
result, can be selected afterwards.
Display:
mode
M6: Averaging
Measurements: 123456789
The number of the single measurements can selected using the
buttons
and the relevant individual measurement can be activated (number is displayed) or
deactivated (underscore is displayed) using the
+
keys for averaging.
For example: From 5 single measurements, measurements 2, 3 and 4 should be
used to calculate an average value.
Display:
mode
Averaging
Measurements: _234_