FCU 8000 series
Switching on the FCU
HYDAC FILTER SYSTEMS GMBH
en(us)
Page 39/112
BEWA FCU8000 3109776d en-us 2011-06-10.doc
2011-06-10
Switching on the FCU
Switching on the FCU
When the FCU is switched on, various things are shown in the display, i.e. model
no., firmware version, available memory, battery charge status, bus address and,
possibly, error messages (cf. Error Messages / Troubleshooting chapter).
Example:
1.
HYDAC FILTER SYSTEMS
FCU 8110 V3.22
2.
Memory: 25.4% 61
Battery: 13.49 V 80%
3.
Bus address: 1
Battery: 13.49 V 80%
4.
Viscosity range:
1 … 10 mm²/s
5.
Measure -> OK/START
Abort -> STOP
The FCU is ready for operation.
Selecting the Viscosity Range
During startup, the viscosity range currently set is shown in the display. Before the
first measurement, check the viscosity of the fluid to be measured and correct the
settings if required. You can find details on change-over of the viscosity range on
page 82.
The FCU provides two viscosity ranges for selection:
1 … 10 mm²/s
5 … 1000 mm²/s