-------------------------------------------------------------------------------------------------- Non-Contact Tonometer HNT-1/1P 47
④
IOP MSR SETUP
: iop measure setting value
⑤
TRACK MOTOR SETUP
: track motor setting value
⑥
TARGET DIST SETUP
: target distance setting value
⑦
P MSR SETUP
: pachy measure setting value
⑧
TOUCH CAL
: touch calibration points
4-1-15. Move to Packing Pos
[Purpose
]
Move the head and chin rest to the packing position.
[Execution]
(1)
Select the
“Move to Packing Pos” button at Engineer Mode.
(2)
Then the head and chin rest will automatically move to the packing position.
4-1-16. Save factory settings
[Purpose
]
Save current values of engineer settings.
[Execution]
(1)
Select the
“Save factory settings” button at Engineer Mode.
(2)
Then the current values of settings will be saved.
4-1-17. Restore factory settings
[Purpose
]
Return the settings to the backuped settings.
[Execution]
(1)
Select the
“Restore factory settings” button at Engineer Mode.
(2)
Then all the settings will be saved as backuped settings.