32 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
i.
The “+” button is incremented by 1.
ii.
The “-“ button is decremented by 1.
iii.
The “+++” button is quickly incremented by 10(current) or 50(time).
iv.
The “+++” button is quickly decremented by 10(current) or 50(time).
(6)
In the graphic mode, you can see the pressure results of the parameters set
through the test.
i.
start
: solenoid measure start time
ii.
pres_idx
: pressure index
iii.
max val
: maximum pressure value
iv.
appl_idx1
: 1st peak place of graph
v.
val
: 1st peak value of graph
vi.
pres1
: pressure value on 1st peak
vii.
appl_idx2
: 2st peak place of graph
viii.
val
: 2st peak value of graph
ix.
pres2
: pressure value on 2st peak
(7)
Press
“Default” button to reset all changes to default settings.
(8)
Press
“Save” button to save the current settings.
4-1-4. Focus Led Setup
[Purpose]
Adjust the LED brightness to find the target effectively.