-------------------------------------------------------------------------------------------------- Non-Contact Tonometer HNT-1/1P 31
[Menu]
(1)
SEARCH OBS CAM WHITE DEAD PIXEL(HIDE)
- Search for white dead pixels in OBS cameras and display their location.
(2)
SEARCH OBS CAM BLACK DEAD PIXEL(OPEN)
- Search for black dead pixels in OBS cameras and display their location.
(3)
SHOW OBS CAM DEAD PIXEL
- Displays the position of all dead pixels of the OBS camera on the screen.
4-1-3. Solenoid Control
[Purpose]
Control the gain of Internal Pressure sensor, for the better measuring graph of
pressure.
[Mode]
(1)
Text
: Set power and timing of solenoid.
(2)
Graphic
: Display the pressure graph after parameter settings.
[Setup Procedure]
(1)
Enter the Solenoid Control Mode at Engineer Mode.
(2)
Select “m30” or “m60” mode.
(3)
Each mode can set seven parameters for current and time of solenoid.
(4)
To set the parameter, press the corresponding parameter.
(5)
Use “+”, “-“, “+++”, “---“ buttons to change the value.