34 Non-Contact Tonometer HNT-1/1P ------------------------------------------------------------------------------------------------
[Preparation]
(1)
IOP Model eyes(IOP 20, 30, 40, 50, 60) in SETUP kit
(2)
Stand Model Eye for each shooting mode.
[Setup Procedure]
(1)
Enter the IOP Measure Setup Mode at Engineer Mode
(2)
There are 4 modes(SPC30, M30, SPC60, M60) and IOP offset settings.
(3)
Press the table space to start IOP measurement setup.
(4)
Then set the target IOP of the current model eye.
-
Adjust target IOP with “+”, “-“
(5)
Press the
“Start” button to automatically track the target.
(6)
Press the measuring button on the joystick lever.
(7)
Then air puffing will start automatically to count down 5 times, shot the solenoid
and display 5 results.