-------------------------------------------------------------------------------------------------- Non-Contact Tonometer HNT-1/1P 41
(15)
To restore the value of setting, choose
“Clear” on the screen.
(16)
To save the current value of setting, choose “Save” on the screen.
4-1-9. Touch Calibration
[Purpose
]
The touch screen is calibrated for correct touch operation.
[Setup Procedure]
(1)
Enter the Touch Calibration Mode at Engineer Mode or Press joystick button
when the screen shows start logo and boot messages on the screen just after
turn on the device.
(2)
Press
“Start” button on the screen or press joystick button.
(3)
Touch the cross shape shown on the screen. To point exact position, use tools