[Preparation]
(1)
REF Model Eye
(2)
Model Eye Stand
[LED Brightness Adjustment]
(1)
Enter the Focus Led Setup Mode at Engineer Mode
(2)
In case of necessary, change
LED Duty values at below
A.
Pachy LED(only HNT-1P) : Pachymetry LED duty
B.
Fixation LED
: Fixation LED duty
C.
External LED(R)
: Right External LED duty
D.
External LED(L)
: Left External LED duty
E.
Aiming LED
: Aiming LED duty
F.
PD(APPL) LED
: PD LED duty
(3)
Use “+”, “-“, “+++”, “---“ buttons to change the value.
(4)
To restore the initial value of LED duty, Choose
“Clear” on the screen.
(5)
To save the current value of LED duty, Choose “Save” on the screen.
4-1-5. IOP Measure Setup
[Purpose]
Set the table to change internal pressure to IOP.