Taking a Monitored Measurement
To take a monitored temperature (Celsius Monitored or Fahrenheit Monitored indicator
shall be on), the procedures for monitored and predictive temperature measurements are
the same, with the following exceptions:
For monitored measurements:
•
The monitor must be set to take a monitored temperature.
•
The monitor displays the temperature continuously.
•
The measurement continues until the probe is replaced in the probe well.
Note:
Verify that the temperature measurement type is set to monitored.
If the measured temperature remains below the minimum measurable temperature for 5
minutes, the temperature cycle is terminated and the temperature display goes blank. To
reactivate the probe, insert it into the probe well, discarding the used probe cover, and
extract it again, loading a new probe cover.
°F
°C
M
SYS
DIA
MAP
%SpO
2
/min
°F
°C
M
mmHg
kPa
BP
1
SYS
DIA
MAP
%SpO
2
/min
°C
mmHg
°C
M
M
Figure 28. Temperature – Monitored Mode
Note:
The monitor only displays 26°C to 43°C (80°F to 110°F). If the reading is over
43°C (110°F) or under 26°C (80°F), the monitor will display as follows.
°F
°C
M
°F
°C
M
°C
M
°C
M
Figure 29. Example display of over 43°C or under 26°C
Temper
atur
e Monitoring