UltraScan PRO Supplemental Manual for EasyMatch QC Ver. 2.0
23
Figure 26. Haze Standardization: White Tile
Figure 27. Haze Standardization: Light Trap
5.
Read the standard or sample by choosing
MEASUREMENTS > READ STANDARD
or
MEASUREMENTS > READ SAMPLE
, clicking the
READ STANDARD
or
READ SAMPLE
button
on the toolbar, or pressing
F2
or
F3
. The following prompt appears.
6.
Place your sample against the transmission port (the hole in the sphere inside the
transmission compartment) and place the white calibration tile at the reflectance port.
Click
READ
. The instrument reads and then the following prompt appears.
Figure 28. Read Sample with White Tile